The TLC SCANNER II for densitometric measurement of thin-layer chromatograms and other objects up to size 200 x 200 mm.
With built-in option for measurements in transmitted light can transparent objects, eg Electropherograms are evaluated.
The scanner does the actual densitometric measurement function, the processing of measured data is an appropriate peripheral device.
In most cases, is used as such, the CAMAG PC evaluation system, consisting of the CAMAG laboratory data system - IBM-PC compatible computer 386 or higher quality - in conjunction with the CATS software. In this case, is also the control of the scanner by the computer. In the CATS manual using the scanner as part of the CAMAG PC evaluation system is described.
The scanner can be operated with simpler peripherals (integrator) For this purpose, the scanner is equipped with its own programming option.
The present manual deals with the device itself, describes its use with the simpler peripherals, but has repeatedly pointed to the connection with the CAMAG PC evaluation system.
They are taken all precautions to ensure that the use of appropriate standard work rules (SOP = Standard Operating Procedures) under conditions of Good Laboratory Practice (GLP) can work.
TECHNICAL DATA
The light sources D2, W and Hg lamps are installed in the device. Their mechanical switching is done by the lateral adjustment lever. All three lamps are current and stabilized voltage.
Tungsten halogen lamp usable continuum 400-800 nm Hg-high pressure lamp line spectrum 254-578 nm
Deuterium lamp usable continuum 200-400 nm
The lamp supply lamp power supply is built into the scanner. Which is located in the beam lamp automatically when the device is switched on.
Pilot Lamp The image of the slit is illuminated by the independent Monochromator with visible light.
Messabteilbeleuchtung 4 Watt Lamp: 254 or 366 nm or "COOL WHITE"
Apochromatic optical system Suprasil fluorite optics with a band of 195-800 nm switch to MACRO and MICRO-figure.
Monochromator grating monochromator, bandwidth switchable 10/30 nm, wavelength range 200-800 nm stepper motor drive, accuracy 1 nm
Measuring gap Macro 0.25 - 6 mm. Column length in seven steps from 0.5 - 12 mm adjustable.
MICRO 0025 - 0.6 mm. Column length in seven steps from 0.05 - 1.2 adjustable.
Photomultiplier detector broadband, multi-alkali type, spectral-sensitivity receiver 185 -800 nm = 0.2 ns response time.
Mesasure-table-movement (Drive): For each axis, two independent
Stepper motors. Reproducibility 0.1 mm.
Mesasure-table-movement manually:
with hotkey: 20 mm / s, and 20 deg / s.
without shortcut: 1 mm / s, or 1 degree / s.
Selectable measuring 0.1 - 4 mm / s in 0.1 mm increments, or 0.1 degrees / s
in 0.1 degree increments. Return under program control 20 mm / s.
While using the evaluation software CATS are 0.1 to 20 mm / s measurement possible.
Power supply 100 V, 120 V, 220 V and 240 V switchable, 50/60 Hz
Power consumption 250 W
Nitrogen about 0.5 bar, 6 L / min
Analog output 0 - 100 mV
Control by Z 80 (2.5 MHz)
Outputs for external output to control the recorder and the control spring feed withdrawal (open collector TTL).
Output for control of peripherals (eg integrator) open collector TTL.
Other Accessories: Output RS 232 for PC CAMAG evaluation system.
Entrance to the interruption of the automatic measuring process.
Interface Built for HP 3390 A HP 3394 A, SP 4100, SP 4200, SP 4270/90 MERCK-HITACHI D-2000/D-2500. HP 3396 A (requires firmware version 3.15)
RS 232 for PC CAMAG evaluation system (comes with the same included). A Scanner firmware version 3.14 is required.
Weight: 38 kg
Dimensions Width = 570 mm, Depth = 735 mm, height = 360 mm
More pictures available on request - thank you!
Handbook, PC-System and Software